Publication: Simple in-house flow-injection capillary electrophoresis with capacitively coupled contactless conductivity method for the determination of colistin
0
0
Issued Date
2015
Resource Type
File Type
application/pdf
ISSN
16159306
Other identifier(s)
2-s2.0-84925349843
Rights Holder(s)
มหาวิทยาลัยศรีนครินทรวิโรฒ
Bibliographic Citation
Journal of Separation Science. Vol 38, No.6 (2015), p.1035-1041
Suggested Citation
Chaisuwan P., Moonta T., Sangcakul A., Nacapricha D., Wilairat P., Uraisin K. Simple in-house flow-injection capillary electrophoresis with capacitively coupled contactless conductivity method for the determination of colistin. Journal of Separation Science. Vol 38, No.6 (2015), p.1035-1041. doi:10.1002/jssc.201401101 Retrieved from: https://hdl.handle.net/20.500.14740/6168
Abstract
An in-house flow-injection capillary electrophoresis with capacitively coupled contactless conductivity detection method was developed for the direct measurement of colistin in pharmaceutical samples. The flow injection and capillary electrophoresis systems are connected by an acrylic interface. Capillary electrophoresis separation is achieved within 2 min using a background electrolyte solution of 5 mM 2-morpholinoethanesulfonic acid and 5 mM histidine (pH 6). The flow-injection section allows for convenient filling of the capillary and sample introduction without the use of a pressure/vacuum manifold. Capacitively coupled contactless conductivity detection is employed since colistin has no chromophore but is cationic at pH 6. Calibration curve is linear from 20 to 150 mg/L, with a correlation coefficient (r2) of 0.997. The limit of quantitation is 20 mg/L. The developed method provides precision, simplicity, and short analysis time. © 2015 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.
Subject(s)
Amino acids
Chromophores
Back ground electrolyte
Capacitively coupled contactless conductivity detection
Capillary electrophoresis separation
Colistin
Correlation coefficient
Flow injection
Limit of quantitations
Polymyxin E
Capillary electrophoresis
2 morpholinoethanesulfonic acid
Amino acid
Colistin
Histidine
Unclassified drug
Antiinfective agent
Colistin
Article
Capillary electrophoresis
Chromatophore
Conductance
Drug determination
Flow injection capillary electrophoresis
Hydrodynamics
Limit of quantitation
Membrane filter
PH
Priority journal
Capillary electrophoresis
Devices
Evaluation study
Flow injection analysis
Procedures
Anti-Bacterial Agents
Colistin
Electrophoresis, Capillary
Flow Injection Analysis
Chromophores
Back ground electrolyte
Capacitively coupled contactless conductivity detection
Capillary electrophoresis separation
Colistin
Correlation coefficient
Flow injection
Limit of quantitations
Polymyxin E
Capillary electrophoresis
2 morpholinoethanesulfonic acid
Amino acid
Colistin
Histidine
Unclassified drug
Antiinfective agent
Colistin
Article
Capillary electrophoresis
Chromatophore
Conductance
Drug determination
Flow injection capillary electrophoresis
Hydrodynamics
Limit of quantitation
Membrane filter
PH
Priority journal
Capillary electrophoresis
Devices
Evaluation study
Flow injection analysis
Procedures
Anti-Bacterial Agents
Colistin
Electrophoresis, Capillary
Flow Injection Analysis