Disconnected et Tube Detection System in Pressure-cycled Ventilator

dc.contributor.authorSueaseenak D.
dc.contributor.authorDuangsasidhorn B.
dc.contributor.authorPrompakorn M.
dc.contributor.authorTaweerutchana R.
dc.contributor.authorTulatamakit S.
dc.contributor.authorSanpinich A.
dc.date.accessioned2021-04-05T03:01:35Z
dc.date.available2021-04-05T03:01:35Z
dc.date.issued2020
dc.date.issuedBE2563
dc.description.abstractThis paper presents the development of a disconnected endotracheal tube detection device. The M5stack development kit and MPXV5050GP pressure sensor were used in this project. The test is including both detected pressure accuracy and disconnected tube detection accuracy. The proposed method consists of prototype design and creating detection algorithms. The performance testing of the system by comparing it to iVent201 ventilator and calibration manometer. The experiment results of our prototype show that the mean accuracy of both pressure value and disconnected tube detection is above 87 percent. © 2020 IEEE.
dc.format.mimetypeapplication/pdf
dc.identifier.citationLifeTech 2020 - 2020 IEEE 2nd Global Conference on Life Sciences and Technologies. (2020), p.189-192
dc.identifier.doi10.1109/LifeTech48969.2020.1570619242
dc.identifier.other2-s2.0-85085194145
dc.identifier.urihttps://hdl.handle.net/20.500.14740/4637
dc.rights.holderScopus
dc.subject.otherDetection accuracy
dc.subject.otherDetection algorithm
dc.subject.otherDetection device
dc.subject.otherDetection system
dc.subject.otherPerformance testing
dc.subject.otherPressure values
dc.subject.otherPrototype designs
dc.titleDisconnected et Tube Detection System in Pressure-cycled Ventilator
dc.typeConference Paper