Publication:
Mobile Sensor Application for Kinematic Detection of the Knees

dc.contributor.authorTossaphon Jaysrichai
dc.contributor.authorAreerat Suputtitada
dc.contributor.authorWatcharapong Khovidhungij
dc.contributor.orgunitคณะกายภาพบำบัด
dc.date.accessioned2022-09-07T08:17:43Z
dc.date.available2022-09-07T08:17:43Z
dc.date.issued2015
dc.date.issuedBE2558
dc.description.abstractObjective To correctly measure the knee joint angle, this study utilized a Qualisys motion capture system and also used it as the reference to assess the validity of the study's Inertial Measurement Unit (IMU) system that consisted of four IMU sensors and the Knee Angle Recorder software. The validity was evaluated by the root mean square (RMS) of different angles and the intraclass correlation coefficient (ICC) values between the Qualisys system and the IMU system. Methods Four functional knee movement tests for ten healthy participants were investigated, which were the knee flexion test, the hip and knee flexion test, the forward step test and the leg abduction test, and the walking test. Results The outcomes of the knee flexion test, the hip and knee flexion test, the forward step test, and the walking test showed that the RMS of different angles were less than 6°. The ICC values were in the range of 0.84 to 0.99. However, the leg abduction test showed a poor correlation in the measurement of the knee abduction-adduction movement. Conclusion The IMU system used in this study is a new good method to measure the knee flexion-extension movement.
dc.format.mimetypeapplication/pdf
dc.identifier.citationAnn Rehabil Med 2015;39(4):599-608
dc.identifier.doihttps://doi.org/10.5535/arm.2015.39.4.599
dc.identifier.urihttps://hdl.handle.net/20.500.14740/11797
dc.language.isoeng
dc.rightsผลงานนี้เผยแพร่ภายใต้ สัญญาอนุญาตครีเอทีฟคอมมอนส์แบบ แสดงที่มา-ไม่ใช้เพื่อการค้า-ไม่ดัดแปลง 4.0 (CC BY-NC-ND 4.0)
dc.rights.holderมหาวิทยาลัยศรีนครินทรวิโรฒ
dc.subject.otherKnee joint
dc.subject.otherMotion
dc.subject.otherValidation studies
dc.subject.otherSensors
dc.subject.otherWireless technology
dc.titleMobile Sensor Application for Kinematic Detection of the Knees
dc.typeArticle
dcterms.accessRightsopen access
dspace.entity.typePublication
mods.urlhttps://www.e-arm.org/journal/view.php?number=570