Publication: A study on volumetric magnetic susceptibility of weight due to its manufacturing process
0
0
Issued Date
2010
Resource Type
File Type
application/pdf
Other identifier(s)
2-s2.0-84872587711
Rights Holder(s)
มหาวิทยาลัยศรีนครินทรวิโรฒ
Bibliographic Citation
21st Conference on Measurement of Force, Mass and Torque Together with HARDMEKO 2010 and 2nd Meeting on Vibration Measurement, IMEKO TC3, TC5 and TC22 Conferences. Vol , No. (2010), p.303-306
Suggested Citation
Sukhon R., Kongchana R., Khongsiri K., Puntaratronnugoon B., Tulasombut V. A study on volumetric magnetic susceptibility of weight due to its manufacturing process. 21st Conference on Measurement of Force, Mass and Torque Together with HARDMEKO 2010 and 2nd Meeting on Vibration Measurement, IMEKO TC3, TC5 and TC22 Conferences. Vol , No. (2010), p.303-306. Retrieved from: https://hdl.handle.net/20.500.14740/7449
Abstract
The OIML R 111-1 requirements on magnetic properties such as the volumetric magnetic susceptibility (χ) and the magnetization of weights used in legal metrology. So far, these requirements cannot met by weights produced locally in Thailand. This research, therefore, aims to study manufacturing processes that may influence the magnetic properties. In order to reduce the magnetic susceptibility of the weights. Preliminary results show that the magnetic properties are stainless steel type dependent. Moreover, controlling process such as cutting, turning with lathe, grinding and heat treatment process can reduce the volume magnetic susceptibility of weights.