dc.date.issuedBE2563
dc.description.abstractA disposable screen-printed carbon electrode (SPCE) modified with an ionic liquid/graphene composite (IL/G) exhibits a wider potential window, excellent conductivity, and specific surface area for the improvement in the voltammetric signal of rapamycin detection. The modified composite was characterized by transmission electron microscopy (TEM), scanning electron microscopy (SEM), and electrochemical impedance spectroscopy (EIS). The electrochemical behavior of rapamycin at the modified SPCE was investigated by cyclic and square wave voltammetry in 60:40 EtOH: 0.1 M LiClO4 at pH 5.0. A high reproducible and well-defined peak with a high peak current were obtained for rapamycin detection at a position potential of + 0.98 V versus Ag/AgCl. Under the optimized conditions, the rapamycin concentration in the range 0.1 to 100 μM (R2 = 0.9986) had a good linear relation with the peak current. The detection limit of this method was 0.03 μM (3SD/slope). The proposed device can selectively detect rapamycin in the presence of commonly interfering compounds. Finally, the proposed method was successfully applied to determine rapamycin in urine and blood samples with excellent recoveries. These devices are disposable and cost-effective and might be used as an alternative tool for detecting rapamycin in biological samples and other biological compounds. [Figure not available: see fulltext.] © 2020, Springer-Verlag GmbH Austria, part of Springer Nature.
dc.format.mimetypeapplication/pdf
dc.identifier.citationMicrochimica Acta. Vol 187, No.4 (2020)
dc.identifier.doi10.1007/s00604-020-4190-2
dc.identifier.issn263672
dc.identifier.other2-s2.0-85082380838
dc.identifier.urihttps://hdl.handle.net/20.500.14740/4618
dc.rights.holderScopus
dc.subject.otherGraphite
dc.subject.otherIonic liquid
dc.subject.otherRapamycin
dc.subject.otherBlood
dc.subject.otherDevices
dc.subject.otherElectrochemical analysis
dc.subject.otherElectrode
dc.subject.otherLimit of detection
dc.subject.otherProcedures
dc.subject.otherUrine
dc.subject.otherElectrochemical Techniques
dc.subject.otherElectrodes
dc.subject.otherGraphite
dc.subject.otherIonic Liquids
dc.subject.otherLimit of Detection
dc.subject.otherSirolimus
dc.titleWide electrochemical window of screen-printed electrode for determination of rapamycin using ionic liquid/graphene composites
dc.typeArticle
dspace.entity.typePublication
swu.datasource.scopushttps://www.scopus.com/inward/record.uri?eid=2-s2.0-85082380838&doi=10.1007%2fs00604-020-4190-2&partnerID=40&md5=33bcd02998f3a750890a48f8b8156342

Files