Title: Analysis of Current Patterns to Determine the Bladder Volume by Electrical Impedance Tomography (EIT)
Authors: Duongthipthewa O.
Uliss P.
Pattarasritanawong P.
Sukaimod P.
Ouypornkochagorn T.
Keywords: Biomedical engineering
Electric impedance
Electric impedance measurement
Electric impedance tomography
Electrodes
Noninvasive medical procedures
Adjacent patterns
Conductivity changes
Electrical impe dance tomography (EIT)
Noninvasive methods
Plane configurations
Reconstruction image
Strong correlation
Voltage response
Image reconstruction
Issue Date: 2020
Abstract: A non-invasive method based on Electrical Impedance Tomography (EIT) is represented here to determine the bladder volume which indicates the volume of urine. Two different current patterns using with EIT i.e. opposite pattern and adjacent pattern (which is more popular) were investigated by simulation on six 3D-realistic models including bone, rectum, and bladder. Sixteen electrodes were attached by two-plane configuration. Five geometries of the bladder having different volumes were used to determine the boundary voltage response to the conductivity change. A case of the bladder volumes was used to determine the performance of image reconstruction of these patterns. The simulation result shows a strong correlation between the voltage responses and the volume changes, in inverse manner. Using the opposite pattern conducted larger amplitude voltage response than using the adjacent pattern, and resulted in higher sensitivity in the upper mid-region between the electrode planes. The reconstruction image of both patterns can clearly locate the bladder. The reconstructed bladder in the case of using the opposite pattern is larger in amplitude than using the adjacent pattern. However, the image of the opposite pattern has more artifacts at the boundary. This is probably due to the incomparable number of current patterns and measurements of each pattern, and the insufficiency of the current flow patterns of the opposite pattern designed in this study. The higher sensitivity to noise and error of the opposite pattern is probably another cause as well. © 2020 ACM.
URI: https://ir.swu.ac.th/jspui/handle/123456789/11836
https://www.scopus.com/inward/record.uri?eid=2-s2.0-85088163062&doi=10.1145%2f3397391.3397433&partnerID=40&md5=c6e071574a5efd38598b72cac4f79da5
Appears in Collections:Scopus 1983-2021

Files in This Item:
There are no files associated with this item.


Items in SWU repository are protected by copyright, with all rights reserved, unless otherwise indicated.