Title: | A Multiple processing hybrid flow system for analysis of formaldehyde contamination in food |
Authors: | Teerasong S. Amornthammarong N. Grudpan K. Teshima N. Sakai T. Nacapricha D. Ratanawimarnwong N. |
Issue Date: | 2010 |
Abstract: | This work proposes a flow system suitable for the rapid screening of formaldehyde contaminated in food. The system is based on the concept of a flow analyzer with a Hantzsch reaction. An operating procedure was developed for multiple tasking and high sample throughput. This resulted in a significant sample throughput of 51 samples h -1 . Under the optimized conditions, linear calibration from 10 to 100 μM was obtained. The system gave a limit of detection and a limit of quantitation of 0.06 and 0.10 mg kg -1 , respectively. The system was successfully applied to re-hydrated dry squids, vegetables and mushrooms. 2010 © The Japan Society for Analytical Chemistry. |
URI: | https://ir.swu.ac.th/jspui/handle/123456789/14753 https://www.scopus.com/inward/record.uri?eid=2-s2.0-77955723296&doi=10.2116%2fanalsci.26.629&partnerID=40&md5=d227c03a22bffdb34b80688c401f8d2c |
ISSN: | 9106340 |
Appears in Collections: | Scopus 1983-2021 |