Keywords: | acetic acid derivative BSS solution eye drops mineral sodium chloride viscoelastic substance article body temperature chemistry comparative study computer simulation cornea endothelium drug combination finite element analysis heat human metabolism phacoemulsification Acetates Body Temperature Computer Simulation Drug Combinations Endothelium, Corneal Finite Element Analysis Hot Temperature Humans Minerals Ophthalmic Solutions Phacoemulsification Sodium Chloride Viscoelastic Substances |
วันที่เผยแพร่: | 2010 |
บทคัดย่อ: | PURPOSE: The purpose of this study was to estimate and compare the temperatures of two different anterior chamber solutions at the corneal endothelial level during phacoemulsification. An ophthalmic viscosurgical device (OVD) and balanced salt solution (BSS) were compared using the finite element method (FEM). MATERIALS AND METHODS: The thermal properties of an OVD (IAL-F) and BSS were studied in an experimental setting. A computer-aided design model of ocular anatomy was created in two dimensions. The phaco needle was considered to be the only source of heat generation. Then, the FEM was used to demonstrate the transient temperature distribution in the two ocular models at 10, 20, 30, 40, 50 and 60 s. In these models, the anterior chamber was filled with IAL-F (IAL-F model) or BSS (BSS model). RESULTS: The heat generation rate of the phaco needle was 0.0004 cal/s/mm(2). The maximum corneal endothelial temperatures for the two models at 60 s were 52.67 and 41.57 degrees C, respectively. The experimental IAL-F model showed fewer changes in temperature for any given time and location. At larger distances from the heat source, less temperature variation was detected. CONCLUSION: Phacoemulsification is a potential heat-generating procedure performed between the delicate anterior chamber structures. During this procedure, IAL-F protects the endothelium against heat better than BSS. Copyright 2009 S. Karger AG, Basel. |
URI: | https://ir.swu.ac.th/jspui/handle/123456789/14745 https://www.scopus.com/inward/record.uri?eid=2-s2.0-73649122497&doi=10.1159%2f000272020&partnerID=40&md5=980280b55570ade8a18b394772be3a0e |
ISSN: | 14230259 |
Appears in Collections: | Scopus 1983-2021 |
Files in This Item:
There are no files associated with this item.
Items in SWU repository are protected by copyright, with all rights reserved, unless otherwise indicated.