ชื่อเรื่อง: | Direct injection of human serum and pharmaceutical formulations for glucosamine determination by CE-C 4D method |
ผู้แต่ง: | Chaisuwan P. Kongprasertsak T. Sangcakul A. Smith N.W. Nachapricha D. Wilairat P. Uraisin K. |
Keywords: | Back ground electrolyte Capacitively coupled contactless conductivity detection Detection limits HPLC method Human serum matrix Matrix removal Migration time Peak area Pharmaceutical formulation Protein precipitation Protonated Serum samples Body fluids Capillary electrophoresis Chromatography Detectors Glucosamine High performance liquid chromatography Precipitation (chemical) Direct injection electrolyte glucosamine article capillary electrophoresis capillary electrophoresis capacitively coupled contactless conductivity detection carbohydrate analysis conductance drug formulation high performance liquid chromatography priority journal proton transport serum Chemistry, Pharmaceutical Electrophoresis, Capillary Glucosamine Humans Pharmaceutical Preparations Serum |
วันที่เผยแพร่: | 2011 |
บทคัดย่อ: | A simple CE-C 4D method has been developed for the determination of glucosamine by direct injection of human serum and pharmaceutical samples. Glucosamine was electrokinetically injected and analysed in its protonated form using 20mM MES/His (pH 6) as background electrolyte in order to separate it from the matrix and to provide a better response to the C 4D detector. Separation of glucosamine in human serum and pharmaceutical samples was performed in 3min without the need for protein precipitation or matrix removal. Good precision in terms of %RSD for the migration time and peak area were less than 1.91% (n=10). The conductivity signal was linear with glucosamine concentration in the range 0.10-2.50mg/mL, with a detection limit of 0.03mg/mL. Recoveries of glucosamine in serum and pharmaceutical samples were 86.5-104.78%. The method was successfully applied for the determination of the glucosamine content in pharmaceutical formulations and validated with high performance liquid chromatography (HPLC). Good agreements were observed between the developed method, label values and the HPLC method. Glucosamine could be detected in spiked serum sample by direct injection. This was not possible by HPLC due to co-eluting interferences. © 2011 Elsevier B.V. |
URI: | https://ir.swu.ac.th/jspui/handle/123456789/14490 https://www.scopus.com/inward/record.uri?eid=2-s2.0-79960560178&doi=10.1016%2fj.jchromb.2011.05.045&partnerID=40&md5=ca2e302c11da5ded412a6b8ba45f9687 |
ISSN: | 15700232 |
Appears in Collections: | Scopus 1983-2021 |
Files in This Item:
There are no files associated with this item.
Items in SWU repository are protected by copyright, with all rights reserved, unless otherwise indicated.