ชื่อเรื่อง: Effect of hoop stress fracture on micromotion of textured ingrowth stems for radial head replacement
ผู้แต่ง: Chanlalit C.
Shukla D.R.
Fitzsimmons J.S.
An K.-N.
O'Driscoll S.W.
Keywords: arm prosthesis
article
cerclage
clinical article
human
human tissue
micromotion
motion
priority journal
radius fracture
stress fracture
Aged, 80 and over
Analysis of Variance
Arthroplasty, Replacement, Elbow
Biomechanics
Cadaver
Coated Materials, Biocompatible
Elbow Joint
Elbow Prosthesis
Female
Fractures, Stress
Humans
Male
Motion
Osseointegration
Prosthesis Design
Prosthesis Failure
Radius
Random Allocation
Stress, Mechanical
Surface Properties
วันที่เผยแพร่: 2012
บทคัดย่อ: Background: Successful bone ingrowth around cementless implants requires adequate initial stability. Hoop stress fractures during stem insertion can potentially hinder prosthesis stability. Hypothesis: We hypothesized that an oversized radial head prosthetic stem (1 mm "too large" and causing a hoop stress fracture during insertion) would result in an unacceptable amount of micromotion. Materials and methods: Grit-blasted radial head prosthetic stems were implanted into cadaveric radii. Rasp and stem insertion energies were measured, along with micromotion at the stem tip. The sizes were increased until a fracture developed in the radial neck. Results: Prosthetic radial head stems that were oversized by 1 mm caused small cracks in the radial neck. Micromotion of oversized stems (42 ± 7 μm) was within the threshold conducive for bone ingrowth (<100 μm) and not significantly different from that for the maximum sized stems (50 ± 12 μm) (P ≥.4). Discussion: Contrary to our hypothesis, hoop stress fractures caused by implantation of a stem oversized by 1 mm did not result in loss of stability. Stem micromotion remained within the range for bone ingrowth and was not significantly diminished after the fracture. This suggests that if a crack occurs during the final stages of stem insertion, it may be acceptable to leave the stem in place without adding a cerclage wire. Conclusion: A small radial neck fracture occurring during insertion of a radial head prosthetic stem oversized by 1 mm does not necessarily compromise initial stability. © 2012 Journal of Shoulder and Elbow Surgery Board of Trustees.
URI: https://ir.swu.ac.th/jspui/handle/123456789/14316
https://www.scopus.com/inward/record.uri?eid=2-s2.0-84862651249&doi=10.1016%2fj.jse.2011.05.001&partnerID=40&md5=e96495e951ef512dc67512044192403a
ISSN: 10582746
Appears in Collections:Scopus 1983-2021

Files in This Item:
There are no files associated with this item.


Items in SWU repository are protected by copyright, with all rights reserved, unless otherwise indicated.