ชื่อเรื่อง: Measurement accuracy of total cell volume by automated dialyzer reprocessing: A prospective cohort study
ผู้แต่ง: Kreepala C.
Sangpanich A.
Boonchoo P.
Rungsrithananon W.
Keywords: adult
Article
cellular parameters
clinical article
clinical effectiveness
controlled study
dialysis membrane
dialyzer
dialyzer reprocessing machine
female
high flux dialyzer
human
low flux dialyzer
male
measurement accuracy
middle aged
priority journal
prospective study
total cell volume
volumetry
weight
วันที่เผยแพร่: 2017
บทคัดย่อ: Introduction Dialyzer reprocessing machines have replaced human labor in preparing re-usable dialyzers. It also made the process of total cell volume (TCV) measurement become faster. Nevertheless, there has been a lack of data on efficacy of weight evaluation on TCV by machine compared to volume evaluation by the conventional method. The aim of this study was to evaluate the efficacy of TCV measurement performed by Kidney-Kleen® reprocessing machine, produced by MEDITOP Company in Thailand, compared to that of the conventional method. Methods This prospective cohort study was performed during September 2014 to December 2015.The low-flux (N = 101) and high-flux dialyzers (N = 100) were included for TCV evaluation. Reused times were up to 5 in the low-flux and 20 in the high-flux dialyzers. The Bland Altman analysis was used to evaluate value measured by different methods. Results The values measured by weight evaluation (by machine) were higher than those obtained by volumetric evaluation of the conventional method in the low-flux (0.81 ± 0.20%) and high-flux (1.32 ± 0.39%) dialyzers. The correlation of TCV values of the two methods were r = 0.98, p < 0.001 and r = 0.71, p < 0.001 for the low- and high-flux dialyzers. Moreover, there was robust association and agreement between the two methods, confirmed by the Bland-Altman Analysis, which suggested that the values acquired by machine were within the limits of agreement, indicating acceptable accuracy of equipment. Conclusion The approach of measurement differed from that of the conventional method (weight evaluation was used instead of volumetric evaluation), the reprocessing machine could offer accurate results. © 2017 The Author(s)
URI: https://ir.swu.ac.th/jspui/handle/123456789/13078
https://www.scopus.com/inward/record.uri?eid=2-s2.0-85019027239&doi=10.1016%2fj.amsu.2017.04.019&partnerID=40&md5=7c09765aedb74b94d0cb2df1f7ed4a07
ISSN: 20490801
Appears in Collections:Scopus 1983-2021

Files in This Item:
There are no files associated with this item.


Items in SWU repository are protected by copyright, with all rights reserved, unless otherwise indicated.