DSpace Repository

Effects of miniscrew location on performance of distalization appliance (iPANDA)

Show simple item record

dc.contributor.author Sukjamsri C.
dc.contributor.author Suzuki B.
dc.contributor.author Arun S.
dc.contributor.author Suzuki E.Y.
dc.date.accessioned 2021-04-05T03:22:25Z
dc.date.available 2021-04-05T03:22:25Z
dc.date.issued 2017
dc.identifier.other 2-s2.0-85015842450
dc.identifier.uri https://ir.swu.ac.th/jspui/handle/123456789/13143
dc.identifier.uri https://www.scopus.com/inward/record.uri?eid=2-s2.0-85015842450&doi=10.1109%2fBMEiCON.2016.7859598&partnerID=40&md5=9158bb7e6af4705f19030aee535a33ff
dc.description.abstract Malocclusion is an incorrect positioning of the teeth when the jaws are closed. Several orthodontic devices have been designed to distalize the molar teeth for malocclusion treatment, one of which is the indirect Palatal Miniscrew Anchorage and Distalization Appliance (iPanda). The important component of the iPanda is the miniscrew implant that is inserted along the midline of the palatal bone. As the miniscrew implant is used as an anchor of the distalization device, miniscrew location are important parameters relating to the success of malocclusion treatment. This study aim to verify the optimum insertion site of the miniscrew implant by using finite element method. A 3D model of the miniscrew was virtually inserted into the anterior, middle and posterior part of the palate. The tension load of 200g per side simulating real orthodontic force was applied to the FE model. Stress distribution in the miniscrew and surrounding bone as well as miniscrew displacements are analyzed. The results suggest that the miniscrew should be inserted at the anterior part of the palate. The findings will be subsequently used for further studies relating to the design of iPanda component. © 2016 IEEE.
dc.subject Biomedical engineering
dc.subject Bone
dc.subject Dentistry
dc.subject 3-d modeling
dc.subject FE model
dc.subject Insertion site
dc.subject Malocclusion
dc.subject Molar teeth
dc.subject Orthodontic forces
dc.subject Tension loads
dc.subject Finite element method
dc.title Effects of miniscrew location on performance of distalization appliance (iPANDA)
dc.type Conference Paper
dc.rights.holder Scopus
dc.identifier.bibliograpycitation BMEiCON 2016 - 9th Biomedical Engineering International Conference. (2017)
dc.identifier.doi 10.1109/BMEiCON.2016.7859598


Files in this item

Files Size Format View

There are no files associated with this item.

This item appears in the following Collection(s)

Show simple item record

Search DSpace


Advanced Search

Browse

My Account

Statistics